Why is absorbance negative?
In scientific experiments and data analysis, absorbance is a common physical quantity, especially widely used in spectral analysis. However, many people are confused by the negative value of absorbance. This article will combine the hot topics and hot content on the Internet in the past 10 days, analyze the reasons for the negative absorbance, and display relevant background information through structured data.
1. Definition and calculation formula of absorbance

Absorbance (A) is a physical quantity that describes the degree of light absorption by a substance. Its calculation formula is:
A=-log10(I/I0)
Among them, I0is the incident light intensity and I is the transmitted light intensity. Normally, I ≤ I0, so A ≥ 0. However, in some special cases, A may have a negative value.
2. Reasons why absorbance is negative
Here are a few common reasons for negative absorbance:
| Reason | Description |
|---|---|
| Instrument calibration issues | The instrument is not calibrated correctly or the baseline is not zeroed, causing deviations in the measured values. |
| Sample scattering or fluorescence | The sample scatters or emits fluorescence, causing the transmitted light intensity to be higher than the incident light intensity (I >I0). |
| Improper selection of reference solution | The absorbance of the reference solution is higher than that of the sample being measured, causing the calculation to be negative. |
| math calculation error | Errors occurred during data processing, such as logarithmic calculation errors. |
3. The correlation between hot topics on the Internet in the past 10 days and absorbance
The following are hot topics and hot content related to absorbance in the past 10 days:
| hot topics | Related content | heat index |
|---|---|---|
| Advances in Spectral Analysis Technology | Calibration methods for new spectroscopic instruments | 85 |
| Laboratory error analysis | How to avoid negative values in absorbance measurements | 78 |
| Fluorescent material research | Effect of Fluorescence on Absorbance Measurements | 92 |
| Data Science Applications | Mathematical processing methods for absorbance data | 65 |
4. How to avoid negative absorbance values
To avoid negative absorbance values, the following measures can be taken:
1.Calibrate the instrument correctly: Make sure the spectrometer or spectrophotometer is calibrated and the baseline is zeroed before taking measurements.
2.Choose an appropriate reference solution: The absorbance of the reference solution should be lower than that of the sample to be tested.
3.Check sample properties: Avoid using highly scattering or fluorescent samples.
4.Validate data calculations: Ensure correct mathematical calculations, especially logarithmic operations.
5. Summary
Negative absorbance values are usually caused by issues with instrument calibration, sample characterization, or data processing. Through scientific methods and rigorous experimental operations, such phenomena can be avoided. At the same time, combined with hot topics across the Internet, it can be seen that the research and application of absorbance are still developing, and there may be more new technologies to solve this problem in the future.